How can the shelf life of medical devices be ensured?


With the outbreak of COVID-19, people have begun to place greater emphasis on health. Previously, we shared some insights into issues related to pharmaceutical packaging—today, let’s also discuss the common-sense knowledge about the "shelf life of medical devices."

The shelf life of a medical device refers to the period during which the final product is guaranteed to perform its intended function properly. Once this shelf life expires, the device may no longer meet its established performance standards or reliably deliver its intended functionality, posing potential risks during use.

So, when it comes to ensuring the shelf life of medical devices, Medical Device Packaging Testing It is also crucial. The shelf life of medical devices is Refers to the period during which a medical device, once formed into its final product, can perform its intended function. The endpoint of the shelf life is the product's expiration date. Beyond this period, the medical device will It may no longer meet the expected performance specifications and functionalities, so it needs to be processed in advance to ensure shelf-life compliance. Medical Device Packaging Testing 。

So, from the perspective of transportation packaging, what factors influence the shelf life of medical devices?

External factors primarily include:

Storage conditions, transportation conditions, production methods, production environment, packaging, and the impact of changes in the sources of raw and auxiliary materials—along with other influencing factors.

Internal factors primarily include:

The inherent properties of individual raw materials/components in medical devices; potential interactions among these raw materials/components within the device; possible interactions between the device's raw materials/components and packaging materials (including storage media, such as preservation solutions for corneal contact lenses, etc.); the impact of manufacturing processes on the raw materials/components and packaging materials used in medical devices; the effects of any radioactive substances contained in the device—and their radioactive decay byproducts—on the original materials/components and packaging materials; and the ability of sterile-packaged products to maintain a robust microbial barrier.

Both internal and external factors can influence the technical performance specifications of medical device products to varying degrees, according to Medical Device Packaging Testing The results show that exceeding the tolerance limits can lead to device failure. It’s important to emphasize that not all medical devices require a specific shelf life. When a particular medical device passes Medical Device Packaging Testing When it is determined that the performance of raw materials and packaging materials does not significantly change over time, a strict shelf-life may not be necessary. However, if a medical device exhibits poor stability or poses an excessively high clinical risk, its shelf life must undergo rigorous validation.

So, what are the different types of validation tests for the shelf life of medical devices?

Typically, the types of validation tests for the shelf life of medical devices can be categorized into two main categories: accelerated stability testing and real-time stability testing.

(1) Accelerated Stability Testing

An accelerated stability test involves placing a product under external stress conditions, examining how the material degrades over time under these stresses, and then using established acceleration factors and their relationship to degradation rates to predict the material's performance under normal storage conditions.

Required It should be noted that accelerated stability testing should not be used to verify the shelf life of medical devices if their raw materials or components are prone to degradation and damage at elevated temperatures.

(2) Real-Time Stability Testing

Real-time stability testing involves placing a product under specified storage conditions until its performance indicators no longer meet the required standards.

In the real-time stability test, The applicant should determine appropriate conditions—such as temperature, humidity, and light exposure—based on the product’s actual production, transportation, and storage scenarios, and conduct product testing at specified time intervals. Given that most regions in China have a subtropical climate, it is recommended to set the temperature and humidity conditions for validation tests as follows: 25°C ± 2°C and 60% RH ± 10% RH.

Real-time stability testing and accelerated stability testing for non-active implantable medical devices should be conducted simultaneously. The results of the real-time stability test serve as direct evidence to validate the product's shelf life. If the results from the accelerated stability test differ from those of the real-time test, the real-time stability test results shall prevail.

What are the categories of verification, testing, and evaluation items for the shelf life of medical devices? The testing items include both product performance testing and packaging system performance testing.

Product performance testing should focus on physical and chemical tests closely tied to the shelf life of medical devices. For medical devices where biocompatibility may change over time, a biological evaluation is required. Where appropriate, packaging seal integrity testing can be used as an alternative to sterility testing.

Systematic inspection of packaging refers to Medical Device Packaging Testing , which primarily include medical device packaging testing items such as package integrity, packaging strength, and microbial barrier performance. As a nationally recognized third-party inspection and testing organization, the center boasts extensive hands-on experience in packaging system performance testing. For Medical Device Packaging Testing The dye penetration method for detecting seal leaks in breathable packaging, visual inspection, and the bubble test for identifying leaks in flexible packaging—among the packaging integrity testing procedures—are all capable of delivering professional testing services. Similarly, for packaging strength tests such as the sealing strength of flexible barrier materials, the resistance of unconstrained packaging to internal pressure failure, and simulated transportation tests, we are equipped with state-of-the-art equipment and advanced technical expertise, ensuring reliable results. Medical Device Packaging and Equipment Testing 。

What are the reference standards for validation testing of shelf life for medical devices?

(1) Basic Requirements and Quality System Standards

ISO 11607-1 "Packaging for Terminally Sterilized Medical Devices – Part 1: Materials, Sterile Barrier Systems, and Packaging System Requirements";

ISO 11607-2 "Requirements for Verification of Forming, Sealing, and Assembly Processes for Final Sterilization Medical Device Packaging – Part 2."

(II) Standard Test Methods for Packaging Systems

1. YY/T 0681 Parts 1–5 of "Test Methods for Sterile Medical Device Packaging";

2. ASTM D 4169 "Performance Testing of Shipping Containers and Systems";

3. ASTM F 1608 "Microbial Penetration Resistance Test for Breathable Packaging Materials."

(III) Packaging Material Standards

YY/T 0698 Relevant content from Sections 2–10 of "Packaging Materials for Final Sterilized Medical Devices."